Upgrading Your Brewer Software to Version 3.74
==============================================

Copyright(C) 1996, SCI-TEC Instruments Inc.

    1503 Fletcher Road
    Saskatoon, Sask. S7M 5S5
    Phone: 306 - 934 - 0101
    Fax: 306 - 978 - 2339


1.  Introduction

This document is intended to explain the major changes to the new 3.74
software to users of the old software.


2.  Changes in Software Setup

The following is a summary of the new procedure for upgrading an old
software installation to version 3.74:

*   Use 374CVT.EXE to convert old instrument constant files into
    3.74 versions.  The files generated are OP_ST.nnn, ICFVAL.nnn,
    DCFVAL.nnn, ZSFVAL.nnn, INSTNOTE.TXT and BREWER.BAT.
    INSTNOTE.TXT contains installation instructions similar to the
    ones written here.

*   Check the contents of OP_ST.nnn and ICFVAL.nnn to make sure
    that your settings are OK.  Most importantly, make sure that
    the data directory entry (the second one in OP_ST.nnn, usually
    C:\BDATA\ or similar) is correct.

*   Create a subdirectory under this directory with your
    instrument number as the name.  (It may already exist there.)
    Copy OP_ST.nnn, ICFVAL.nnn, DCFVAL.nnn, ZSFVAL.nnn, and your
    UV response file into this subdirectory.

*   Create a \BREWER directory and copy MAIN.BAS, *.RTN, and *.FIL
    into it.  Place a copy of OP_ST.nnn in this directory also, and
    rename it to OP_ST.FIL.

*   Copy BREWER.BAT to a directory present in the PATH statement.
    Edit BREWER.BAT to include the full path of the Brewer
    directory.

*   Update AUTOEXEC.BAT, if needed, by adding the line
    CALL BREWER.BAT.

*   Run the software by rebooting the computer or by typing BREWER
    at a DOS prompt.

If this is a new installation of the software, use this procedure
instead:

*   Create a \BDATA directory for your Brewer data, and an
    instrument directory under this one.  Copy OP_ST.nnn,
    ICFVAL.nnn, DCFVAL.nnn, ZSFVAL.nnn, and your UV response file
    into the instrument directory.

*   Create a \BREWER directory and copy MAIN.BAS, *.RTN, and *.FIL
    into it.

*   Edit OP_ST.nnn to ensure that the instrument number entry (the
    first entry) and the data directory entry (the second entry)
    are correct.  Edit ICFVAL.nnn to ensure that the instrument
    type (the 23rd entry) and COM port (the 24th entry) are
    correct.  Editing all other instrument constants may also be
    done now, or can be done with CF and IC once the Brewer
    software is running.

*   Place a copy of OP_ST.nnn into the \BREWER directory, and
    rename it to OP_ST.FIL.

*   Copy BREWER.BAT to a directory present in the PATH statement.
    Edit BREWER.BAT to include the full path of the Brewer
    directory.

*   Update AUTOEXEC.BAT, if needed, by adding the line
    CALL BREWER.BAT.

*   Run the software by rebooting the computer or by typing
    BREWER at a DOS prompt.


3.  Software Reorganization

There is a significant change to the organization of the Brewer
software.  The changes are summarized here:

*   There were eight separate versions of the 3.73 Brewer
    software.  For the 3.74 software release, there are only
    three:  a version for the MKII, one for MKIII, and one for
    MKIV.  NO2/O3 modes are handled internally by the MKIV
    software, and the temperature circuit changes are handled
    internally.

*   As previously mentioned, the instrument constants entries
    have been reorganized.  Several entries have also been added.

*   The four instrument constant files (OP_ST.nnn, ICFVAL.nnn,
    DCFVAL.nnn, and ZSFVAL.nnn), and the UV response file are
    separated by instrument.  A copy of OP_ST.nnn is stored in
    the \BREWER directory as OP_ST.FIL to ensure that the
    software knows which instrument number to use upon startup.
    All five of these files are stored in the Brewer data
    directory under each instrument number.  When NO is used to
    switch, the software uses the new instrument number's
    constant files instead.

*   The old software required a MODE.BAS, NOMODE.BAS, and
    DISK.FIL to work properly.  This has been eliminated in the
    new software.


4.  Software Improvements and Changes

The Brewer software has changed significantly since the 3.73 version.
The following subsections discuss these changes.  Refer to the README
file included with this software for any changes since the release of
3.74 (such as 3.74a, 3.74b, etc.)

4.1.    New Brewer Commands

*   In the old software, you had to press CTRL-BREAK before you
    could shell to DOS.  Now, simply type SHELL at the Brewer
    prompt (without pressing CTRL-BREAK) to enter DOS.  Typing
    EXIT will return you to the Brewer program.

*   The old MKIV software used an SW routine to switch operating
    modes.  In 3.74, this has been replaced with N2 and O3.  The
    new commands guarantee that the Brewer ends them in the
    proper operating mode.  Also, SW has been disabled in the
    new software because it is no longer reliable.

4.2.    Changes to Brewer Commands

*   CF now allows you to change the filename of the ICFVAL.nnn
    file.  However, NO must now be used to change the
    instrument number.

*   CI uses the extended range of a MKIII and MKIV Brewer, if
    present.  Also, the scan resolution is selectable to 1, 2,
    or 5 angstroms.

*   The DSP tests now offer use of the extended range of the
    MKIII/MKIV Brewers.  They also offer better (selectable)
    lamp arrangements, and handle variable numbers of scan
    lines for manual entry.

*   Low intensity dead time times were of poor quality in the
    old Brewer software.  DT has been fixed in 3.74 to solve
    this problem.

*   ED and ED-PD have a new command sequence that includes
    all summaries and all standard tests such as DT, RS, and
    HG.

*   EX, INST, and OP_ST are now built into the main program.
    This means that INST and OP_ST do not exist as RTN files
    anymore.

*   FR resets the micrometer(s), and the third filterwheel if
    present, on all Brewer models.  The MR routine has been
    eliminated, and should be replaced with FR in schedules.

*   HG has had many changes to it in the 3.74 software:

    *   HG now has a selectable spectrum, based on
        whether you have a narrow or wide HG exit slit.
        Use IC to set this flag.

    *   HG is disallowed in NO2 mode.

    *   When HG fails, it resets the micrometer(s).  If HG
        fails again, it quits trying and aborts any
        schedule currently running.

*   HP attempts to align the second spectrometer twice.  If it
    fails the first time, it performs an FR.  If it fails the
    second time, it quits and aborts any schedule currently
    running.

*   HP now has an improved algorithm for matching its measured
    spectra.

*   NO will now automatically create a new instrument
    directory if it does not exist, creating new constant
    files from the files currently being used.  The two sets
    of files can now be managed independently.

*   PB now handles the functions of both NPB and OPB, based on
    the current operating mode.

*   PO now prints out all instrument constants.  This means
    that DSPPO is no longer needed.

*   RE now resets every motor, except the micrometer(s) and
    the third filterwheel, on all Brewer models.  This is
    also done when the software is booted.

*   Printout errors in REP, NOSUM, and ZSSUM have been fixed.

*   There is now a single UV and UVSUM routine which are
    capable of handling all types of UV scans:  UA, UF, UV,
    and UX.  All data is written to a single UVdddyy.nnn file.

*   UVSUM had incorrect temperature output values, which have
    now been fixed.

*   The following routines have been removed from the
    software:  ES, ES-PD, FS, HESUM,  SUM_DAY, DC, DIS, DIS1,
    MI, HU, INIT-PF, INIT-PN, SU, SZ, and WS.

*   Several routines missing from the MKIII software have
    been included.  These are AU, AUC, CS, ED-PD, MD, PB, SA,
    and UB.

4.3.    General Bug Fixes

*   On certain months, it is possible to get a day 0 on a
    printout, such as "April 0", in older Brewer software.
    This bug has now been fixed.

*   UA had incorrect timing after switching to extended
    range.  This has been fixed.

*   The Julian day was calculated incorrectly between local
    midnight and solar midnight in certain areas of the
    world, and after the Brewer time was set with TI.  This
    has been fixed.

*   The clock test routine failed at midnight when the clock
    rolled over to the new day.  This has been fixed.

4.4.    General Changes

*   All software routines automatically detect whether they
    are in NO2 or ozone mode.  If you run a routine (on a
    MKIV) from the wrong operating mode, the routine will
    inform you of the fact, and will not run.  Note that
    this does not abort any schedules currently running.

*   The menus have been reorganized to a more intuitive
    structure.

*   The main program, now called MAIN.BAS, is identical for
    all versions of the Brewer software.

4.5.    Changes to Support Utilities

*   LINEFIT.EXE and UMLIN.EXE have been updated to handle
    the new instrument file format.  Also, some known bugs
    were fixed.

*   RD_UX.EXE has been updated to handle the new
    UVdddyy.nnn file format.  Also, it will process raw
    counts into counts/second if no UV response file is
    provided.

4.6.    Changes to Instrument Constant Files

*   As previously mentioned, the instrument constant files
    now consist of ICFVAL.nnn, DCFVAL.nnn, and ZSFVAL.nnn.
    OP_ST.FIL is still present, with a small name change to
    OP_ST.nnn.

*   ICF/DCF/ZSF all contain a release date field, which can
    be used to hold useful reminders about the contents of
    the file, such as a modification date, or a reminder
    that this is a test set of constants.

*   The zenith sky coefficients are now found in the file
    ZSFVAL.nnn.  The 374CVT.EXE conversion utility will
    generate this file, using default values if none are
    provided.

*   The dispersion constants are now found in the file
    DCFVAL.nnn.  The 374CVT.EXE conversion utility will
    generate this file, using the constants from the old
    #nnn.FIL.

*   All other instrument constants are found in the file
    ICFVAL.nnn.  This includes many constants that only
    have meaning for one instrument type.  Any constants
    that fall into this category are ignored if the
    instrument is not the right type.

*   There are several new instrument constants in
    ICFVAL.nnn:

    *   Brewer type (MKII/MKIII/MKIV)
    *   COM port
    *   O3/NO2 positions for filterwheel #1
    *   Standard and UV positions for filterwheel #2


5.  Introduction to the 374CVT.EXE Conversion Utility

The conversion utility will ask you a number of questions about
your old Brewer software and the instrument itself.  Before you
begin, make notes on the following questions:

*   What type of Brewer do you have? (MKII, MKIII, or MKIV)

*   What version of the Brewer software do you have now?
    This number will appear, whenever you run the Brewer
    software, on the title screen.  (Note:  if you have
    version 3.73 or later of the software, then the next
    three questions should already be answered for you.)

*   What micrometer zero position do you use?  This step
    value is normally 3469 on the MKII, 2469 on the MKIII,
    and 2669 on the MKIV.

*   How many steps are required to open the iris?  The old
    string motors need 75 steps, and the new linear drive
    motors need 250 steps.

*   How fast is your computer?  There is a buffer delay
    setting to compensate for fast machines not waiting
    long enough for responses from the Brewer.

*   What positions do you use for filterwheels #1 and #2?
    This information can be found by examining the DT.RTN
    and INIT.RTN files.  In DT.RTN, the variable M4$ is
    set to a specific value, depending on whether you are
    in ozone or NO2 mode (NO2 mode is specific to the MKIV
    Brewer), which is used for positioning the first
    filterwheel.  The variable M5$ is set in DT.RTN as
    well, to specific values in the low intensity and high
    intensity tests, which are used for the second
    filterwheel.  Finally, the variable UF$ (found in
    INIT.RTN) contains the UV mode setting for the second
    filterwheel.  Record all of these values for reference.

*   What zenith sky coefficients do you use?  These values
    can be found in the main program at lines 8900 to 9000.

*   What zenith offset do you need?  This is the distance,
    in steps, from the sensor zeroing position to the hard
    stop.  It is easily measured, and is typically around
    30 to 50 steps.

*   Do you have:

    *   a second film polarizer?
    *   a wide HG exit slit?  (typical of newer Brewers)
    *   new Brewer electronics?  (as of Feb 8/95, these
        are not available to customers yet)

If you have not modified your version of the software, the
default filterwheel positions and zenith sky coefficients
suggested by the program should work.

First, BACK UP your old Brewer software and instrument constants
files.  Copy the following files into a single directory:
#nnn.FIL, OP_ST.FIL, and DISK.FIL.  If you are on a MKIV, you
will also have #000.FIL present in your NO2 directory.  Note
this directory for reference.  Finally, delete all files in your
Brewer and data directories.

You should now install the new Brewer software by copying all
files in A:\BREWER to your Brewer directory.

Now, you are ready to run the conversion utility.


6.  Converting Your Instrument Constant Files

As mentioned above, the old software had three or four constants
files that are now together in one directory.  The conversion
utility will convert these into four files, plus two utility
files.  The new filenames are as follows:

*   ICFVAL.nnn:  This file contains instrument constant
    values.  It is similar to the old #nnn.FIL.

*   ZSFVAL.nnn:  This file contains all zenith sky
    coefficients for your Brewer.

*   DCFVAL.nnn:  This file contains the dispersion
    coefficients for your Brewer.

*   OP_ST.nnn:  This file contains the Brewer's current
    operating state.

*   BREWER.BAT:  This is a sample batch file to demonstate
    the correct method to run the 3.74 versions of the
    Brewer software.

*   NOBREW.BAT:  This is a sample batch file to demonstate
    the correct method to run the 3.74 versions of the
    Brewer software, with no Brewer attached.

*   INSTNOTE.TXT:  This file is a brief summary of the
    installation procedure for the 3.74 software.

You must create a directory to hold these output files, since
the conversion utility will expect you to provide one.

Run the conversion utility by typing 374CVT.  The program will
ask you many questions, based on the information gathered in
Section 2.  Default answers are always given in brackets:  if
you change an answer, the program will re-display it in
brackets to verify.  If an answer is correct, press ENTER
without typing in a new answer.  Be certain to read a question
carefully before typing anything!

Once the program is done, the above six files will be written
to the specified output directory.  The file INSTNOTE.TXT
explains what to do with the output files.

NOTE:  keep a backup copy of your old instrument constant
files for reference.  If you run into trouble with the new
software, you can use CF or IC to manually change the
instrument constant entries.


7.  Troubleshooting

This section attempts to address any problems that you may
run across with the new Brewer software.  If you still have
trouble after reading this section, please contact SCI-TEC
Instruments Inc. with your question.  Please gather as much
information as possible before calling; this makes it much
easier for us to deal with any problem you may have.

*   Problem:  The new Brewer software is really slow!
    It takes forever to accomplish anything!

*   Solution:  If your old software was version 3.73 or
    later, check your original copy of #nnn.FIL.  The
    last entry in the file should be a small number,
    typically in the range 0.2 to 0.8.  This is the
    DELAY setting.  If the last number is much larger,
    such as 75 or 250, then the true last entry is
    missing.  If this is the case, use a text editor to
    add 0.2 to the end of #nnn.FIL.  Then, rerun the
    conversion utility and install the output files as
    before.

*   Problem 1:  The Brewer software can't find any of
    the menus or the RTN files!
*   Problem 2:  The Brewer software always stops with
    the error "missing environment variable BREWDIR".

*   Solution:  You probably have the BREWDIR
    environment variable set wrong (or not at all), or
    you forgot to change to the Brewer directory in
    your batch file.  Make sure that either BREWER.BAT
    or AUTOEXEC.BAT (not both) contains the following
    two commands:

    cd C:\BREWER        ; substitute the Brewer
                  software directory here
    set BREWDIR=C:\BREWER   ; this entry should have
                  exactly the same path as
                  above

*   Problem:  The Brewer software always stops with the
    error "cannot find C:\BDATA-O\OP_ST.000" (or some
    other directory name)!

*   Solution:  Your C:\BDATA-O\nnn\OP_ST.nnn and
    C:\BREWER\OP_ST.FIL files are not identical.  The
    first seven entries in OP_ST.FIL tell the Brewer
    where to find the Brewer data directory, what the
    instrument number is, and what files contain the
    instrument constants.  If any of these entries are
    wrong, you will see the error message.  Fix the
    entries in OP_ST.nnn, copy this new file to your
    \BREWER directory, and rename it to OP_ST.FIL.
    This must be done in the order specified!  If you
    only edit OP_ST.FIL, the error will occur again!

*   Problem:  (MKIV)  Why can't I use the SW command
    anymore?

*   Solution:  The new software does not work with the
    old SW command, so it has been disabled.  Use the
    N2 and O3 commands to switch directly to NO2 or
    ozone mode.  The command will inform you if you
    are already in the mode you requested, but this
    does not hurt anything.

*   Problem:  (MKIV)  Where are the BDATA-N and
    BDATA-O directories?  There used to be separate
    NO2 and ozone mode 'B' files.

*   Solution:  The new MKIV software has been combined
    into one version.  All data goes to a single BDATA
    directory.  The average files contain an entry to
    indicate the current operating mode.  Also, all 'B'
    files have n2 flags in front of data obtained from
    NO2 mode.  For example, NO2 summaries have
    "n2summary" instead of "summary" at the beginning.