UVS-E-T UV Radiometer

  • Spectral response optimised for UVE measurement
  • Excellent directional response
  • UVIATOR software included

Add this product to basket Request information
Share this page

UVS-E-T UV Radiometer

The UVS-E-T measures UVE irradiance to the ISO 17166:1999, CIE S 007/E-1998 response function. The precision quartz dome and diffuser provide class-leading directional response.

This single-band radiometer has one detection system with a 0-3V analogue voltage output.

The instrument is thermostated (controlled at +25 ºC ±2 ºC) including the filters, detector and pre-amplifier.

Due to the heating/cooling control and signal amplifiers, it requires external power.

UVS-E-T has a rugged weather-resistant case made from protected aluminium, is designed for continuous outdoor operation, and can be easily be adjusted to the horizontal plane by three levelling feet and a built-in bubble level.

The screw-in drying cartridge is easy to remove and replacement desiccant is supplied in convenient refill packets.

UVIATOR software
UVS-E-T is supplied with extensive factory calibration data files and our unique UVIATOR software package. These files allow manual correction of the data to minimize errors. However, UVIATOR takes this a step further by carrying out the process automatically and providing graphical visualization of the results.

Combining UVIATOR software with the spectral characteristics of the radiometer in the calibration files, corrects the mismatch between the actual instrument response and the ideal sensor response. This enables further corrections to be made to compensate for changing atmospheric conditions that influence the UV spectrum.

The correction methodology is based on an atmosphere transmission model. UVIATOR calculates the air mass for each stored measurement and retrieves from the internet the appropriate Ozone Monitoring Instrument (OMI) satellite. This satellite gives the total Ozone column density based on position (longitude and latitude) and time.

Using UVIATOR typically reduces the measurement uncertainty by a factor of two.

UVIATOR software is available as a free download from our website.

Back to productcategory

Specifications

Spectral range (overall) ISO 17166:1999, CIE S007/E-1998 nm
Typical output range 0 to 0.6 W/m²
Response time (95%) < 1.8 s
Non-linearity < 1 %
Temperature dependence of sensitivity (-25 °C to +50 °C) no error due to internal temperature stabilization at 25 °C
Operational temperature range -40 °C to +50 °C
Supply voltage DC 7 to 18 V

FAQ