Only once. The calibration is valid for the whole lifetime of the device.
It is needed only once to make sure the effect of your local dust is used to get the best measurements.
The DustIQ does not need recalibration for at least 5-10 years as it has extra internal sensors and compensation for any potential wear or drift.
No. All components have been selected in such a way that we do not anticipate factory recalibration. A local dust calibration also measures and corrects for any future deviation from the original factory calibration.
The DustIQ works with dust it has been calibrated with. For the factory calibration it is Arizona Test Dust. For other dust types, the DustIQ can be calibrated locally, it cannot be calculated or "programmed".
The data will be less accurate most likely. After local dust calibration the "old" data can be post-processed as described in the manual.
Ideally, perform a local dust calibration soon after and record the mentioned Modbus registers for dust slopes before and after. Going back to factory settings is possible.
We have compiled all relevant information on a separate page. Please see: https://www.kippzonen.com/Download/999/DustIQ-Calibration-Procedure-0386910-and-0386915-models